The Welch Allyn 7500 Connex Spot Monitor with WiFi Connectivity includes CSM power supply (#7000-PS), one Adult Cuff, and one Large Adult Cuff. This system comes in several configurations, each containing different built-in features. Please review the configurations tab to see which is right for you.
Warranty: 2 Years.
The Welch Allyn 7500 Connex Spot Monitor with WiFi Connectivity features a user-friendly, bright touchscreen display. The 7500 provides accurate vital signs measurement including blood pressure averaging, spot checking, interval monitoring and custom scoring across patient populations. Its upgradable design and custom configurations include optional wall unit integration as well as multiple thermometry and pulse oximetry choices.
Features
Welch Allyn 7500 Connex Spot Monitor with WiFi Connectivity Features
- SureBP technology can provide blood pressure measurements in only 15 seconds.
- Connex Scoring App calculator for custom protocol.
- Supports resident list workflows for long-term care settings.
- Partnerships with more than 90 leading EMRs including Epic and Cerner.
- Add to integrated wall system for complete patient exam workflow.
- Mobile stand option includes work surface with integrated battery for extended usage.
- Single sign-on options available for quick, secure clinician login.
- BR averaging to help ensure accurate hypertension diagnosis.
- SureTemp Plus or Braun ThermoScan PRO 6000 Thermometry Choice of leading technologies: Masimo, Nellcor, or Nonin.
- Scale connectivity to Detecto, Health-o-Meter, and SECA scales.
Specifications
Welch Allyn 7500 Connex Spot Monitor with WiFi Connectivity Specifications
-
Wireless Specifications:
-
Bluetooth: 2.1 + EDR.
-
Frequency: 2.402 - 2.480 GHz.
-
Max. Transmit Power: Class 1; +8 dBm from Antenna.
-
Receive Sensitivity: -89 dBm.
-
Range: Circa 100 meters.
-
Data Rates: Up to 3 Mbps (over the air).
-
Electrical Rating: 100 - 240 V AC, 50 - 60 Hz, 0.8x - 1.5A.
-
Duty Cycle: Continuous operation.
-
Type of Protection Against Electric Shock: Class I internally powered.
-
Height:
-
Standard Chassis: 2.3" x 9.2" x 6.3" (5.8 cm x 23.4 cm x 16.1 cm).
-
Extended Chassis with Braun: 4.4" x 11.7" x 6.5" (11 cm x 29.8 cm x 16.6 cm).
-
Extended Chassis with SureTemp: 4.2" x 11.7" x 6.4" (10.6 cm x 29.8 cm x 16.6 cm).
-
Weight (including battery):
-
Standard Chassis: 2.9 lb (1.3 kg).
-
Extended Chassis with Braun: 3.7 lb (1.7 kg).
-
Extended Chassis with SureTemp: 3.5 lb (1.6 kg).
-
Graphical Display Resolution:
-
Dimensional Outline: 6.5" x 4.1" x 0.13" (165 mm x 104 mm x 3 mm).
-
Active Area: 6.1" x 3.4" (154 mm x 86 mm).
-
Resolution: 800 x 480 pixels.
-
Pixel Arrangement: RGB.
-
Pixel Size: 63.2 ╬╝m x 179 ╬╝m.
-
Luminance: 530 cd/m2.
-
Speaker Volume:
-
Minimum Output Sound Pressure: 60 dB at 1.0 meter.
-
Alarm and Pulse Tones: per IEC 60601-1-8.
-
Pulse Frequency: 150 - 1000 Hz.
-
Harmonic Components in the Range 300 Hz to 4000 Hz: Minimum of 4.
-
Effective Pulse Duration:
-
High Priority: 75 - 200 ms.
-
Medium and Low Priority: 125 - 250 ms.
-
Noninvasive Blood Pressure:
-
Cuff Inflation Target:
-
Adult: 160 mmHg.
-
Pediatric: 140 mmHg.
-
Neonate: 90 mmHg.
-
Systolic Range:
-
Adult/Pediatric: 30 - 260 mmHg.
-
Neonate: 20 - 120 mmHg.
-
Diastolic Range:
-
Adult/Pediatric: 20 - 220 mmHg.
-
Neonate: 10 - 110 mmHg.
-
MAP Range:
-
Adult/Pediatric: 23 - 230 mmHg.
-
Neonate: 13 - 110 mmHg.
-
Blood Pressure Accuracy: Meets ANSI/AAMI standards for NIBP accuracy (+/- 5 mmHg mean error, 8 mmHg SD).
-
BP Determination Time: 15 seconds typical, 150 seconds max.
-
Overpressure Cutoff:
-
Adult/Pediatric: 300 mmHg +/- 15 mmHg.
-
Neonate: 150 mmHg maximum.
-
Pulse Rate Range:
-
Adult/Pediatric: 30 - 200 bpm.
-
Neonate: 35 - 220 bpm.
-
Pulse Rate Accuracy: +/- 5% (+/- 3 bpm).
Configurations
Welch Allyn 7500 Connex Spot Monitor with WiFi Connectivity Configurations
-
#75MX-B: Connex Spot Monitor with SureBP NIBP and Masimo SpO2.
-
#75CX-B: Connex Spot Monitor with SureBP NIBP and Covidien SpO2.
-
#75CT-B: Connex Spot Monitor with SureBP NIBP, Covidien SpO2, and SureTemp Plus Thermometer.
-
#75MT-B: Masimo SpO2, SureTemp Plus Thermometer.
-
#75CE-B: Connex Spot Monitor with SureBP NIBP, Covidien SpO2, and Braun ThermoScan PRO 6000 Ear Thermometer.
-
#75ME-B: Masimo SpO2, Braun ThermoScan PRO 6000 Ear Thermometer.
Literature
Welch Allyn 7500 Connex Spot Monitor with WiFi Connectivity Literature
Welch Allyn Connex Spot Monitor Brochure
Welch Allyn Connex Spot Monitor Configuration Logic and Ordering
Welch Allyn Connex Vital Signs Monitor with the Connex Scoring App Brochure
Welch Allyn Connex Spot Monitor Service Manual
Welch Allyn Connex Spot Monitor Directions for Use
Customer Reviews 0 out of 5 - (Total 0 ratings)