Refurbishing Process:
All OEC-6800ÔÇÖs go through a multi-step refurbishing process. Each unit is restored to a like-new status. Hand Surgeons, Podiatrists, Orthopedics, Rehabilitation Services are just a few of the applications utilizing the refurbished OEC-6800 C-arm. The Advanced Technology 1Kx1K CCD Camera Technology used in the OEC-6800 image system allows for crisp clear imaging. Whatever the purpose, a refurbished OEC-6800 will deliver the results you can depend on.
Mechanical repairs/cosmetic
maintenance
ÔÇó All defective parts and covers are┬árepaired or replaced
ÔÇó C-arm is completely disassembled┬áand repainted
ÔÇó ÔÇ£CÔÇØ is removed from main frame┬áassembly
ÔÇó Image intensifier removed and┬áinspected for any imperfections┬áand grid integrity
ÔÇó Camera removed, cleaned and tested to OEM Specs┬á
ÔÇó Wheels removed, cleaned and or┬áreplaced from main chassis
ÔÇó Chassis inspected for structural┬ádefects and repaired as necessary
ÔÇó All bearings are inspected, cleaned and re-lubricated, or replaced as needed
ÔÇó Control panel display is tested and┬áreplaced if necessary
ÔÇó All wheels are inspected and┬álubricated and/or replaced as┬áneeded, with new cable pushers┬áinstalled
ÔÇó All locks and brake assemblies are┬áinspected, cleaned, and repaired┬áor replaced as needed
ÔÇó C-arm is completely disassembled┬áand repainted
ÔÇó A complete set of operator and┬áservice manuals are provided
Re-assembly
ÔÇó All components sent to C-arm┬áreassembly area for rebuilding
ÔÇó ÔÇ£CÔÇØ, Image tube and camera,┬áVertical column, main control and┬ápower cable installed
ÔÇó New CRT monitors installed in┬ámonitor cart
ÔÇó Monitor cart rebuilt
ÔÇó When C-arm is completely rebuilt┬ámechanically, system is sent for┬ácalibration and final testing
Testing/calibration
ÔÇó Image intensifier and X-ray tube┬áare tested for stability and balance
ÔÇó Image intensifier is tested for┬áresolution and gain to be within┬áOEM specifications
ÔÇó X-ray tube radiation output is┬áverified to be within OEM┬áspecifications
ÔÇó X-ray generator high voltage tested
ÔÇó Maximum dose rate is set in┬ácompliance with government┬áregulations
ÔÇó CCD camera calibrated to OEM┬áspecifications
ÔÇó X-ray beam is aligned for each┬áfield size
ÔÇó C-arm is tested for current leakage requirements
ÔÇó All monitors are aligned, resized,┬ácentered and focused
ÔÇó All monitors are adjusted for┬ábrightness and linearity
ÔÇó Video system is optimized for gray┬áscale and resolution
ÔÇó All power supplies are tested and┬ácalibrated or replaced as┬ánecessary