Imedical Healthcare Solutions
The QRS Pre-Calibrated Mouthpiece is disposable and calibrated during production using the same equipment LDS hospital uses to validate spirometers against the American Thoracic Society's recommendations for the standardization of spirometry. It is compatible with all QRS spirometers.
Warranty: 90 Days - Free from Defects in Material and Workmanship.
These pneumotachs are individually-wrapped and single patient use only, but can be used up to 20 times over a 10 day period for that patient, if the fabric does not become contaminated. The patient should open, handle, and dispose of his/her own mouthpiece and should return it to its plastic bag to avoid contamination if future use is anticipated.
The QRS mouthpiece reduces the liability for cross-contamination and eliminates the need for sterilization. This can save your staff time and money.
The ATS/ERS 2005 Standardization of Spirometry document recommends a calibration plan for disposable mouthpieces. Because QRS mouthpieces are pre-calibrated in production, there is no need to test every mouthpiece. QRS recommends implementing a quality assurance program to periodically select and test calibration on a sample of disposable mouthpiece pneumotachs.
The Pressure Tube (#Z-7000-2032) connects the Pre-Calibrated Mouthpiece to the QRS spirometers.
QRS Office Medic User's Manual