The GE B105 Patient Monitor includes Printer, Temp, 3 and 5 Lead ECG, Extension Rack, Battery (only compatible with CO2 module), and your choice of SpO2. Some configurations also include Invasive Blood Pressure. Please see the configurations tab to determine your system choice.
Warranty: 1 Year.
The GE B105 Patient Monitor is a portable multi-parameter monitor used for monitoring, recording, and to generate alarms for multiple physiological parameters of patients of all ages in hospital environments and during intra-hospital transport. It is an intuitive, flexible, and reliable system for busy caregivers who need a variety of clinical applications.
The B105 monitors and displays ECG (including ST segment, arrhythmia detection, ECG Diagnostic Analysis and Measurement,), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/Surface temperature, impedance respiration, respiration rate, and CO2.
Features
GE B105 Patient Monitor Features
- Accurate NIBP measurement even in challenging conditions with DINAMAP SuperSTAT NIBP technology, ECG 4 leads simultaneous analysis with EK Pro v14, choice of SpO2 technologies: TruSignal, Masimo SET, Nellcor OxiMax.
- Capacitive touch screen, large numeric mode, choice of numerical or continuous waveforms and care area specific pre-configured workflow settings.
- ST Segment and full Arrhythmia analysis, SpO2, NIBP, IBP, RR, ECG, and EtCO2.
- Records and stores up to 168 hours of monitoring activities across all parameters.
- Analyzes up to 16 types of arrhythmia, including A Fib.
- Three hour battery life.
- Large numeric mode enables critical parameter visibility even up to 13 feet away.
- EK-Pro arrhythmia analysis.
- Connectivity to GE Carescape networks and Mobile Care Web Viewer.
- Flexibility to share parameter modules and accessories across other GE monitors.
- Advanced applications such as ST Segment and Full Arrhythmia.
- Rugged and robust design stands up to harsh environments and everyday wear-and-tear.
Specifications
GE B105 Patient Monitor Specifications
-
Display: 10.1" Display, WVGA.
-
Waveforms: Up to 6 dynamic waveforms.
-
ECG:
-
Leads Available: 3-lead configuration: I, II, III; 5-lead configuration: I, II, III, aVR, aVL, aVF, and V.
-
Sweep Speed: 12.5, 25, or 50 mm/s.
-
Gain Range: 0.5x, 1x, 2x, 4x.
-
Heart Rate Accuracy: 30 to 300 bpm, +/-5% or +/-5 bpm, whichever is greater.
-
Bandwidth:
-
50/60 Hz Power Supply:
-
Monitor: 0.5 to 40 Hz.
-
ST: 0.05 to 40 Hz.
-
Diagnostic: 0.05 to 145 Hz.
-
Pacemaker Detection:
-
Range: 2 to 700 mV.
-
Pulse Width: 0.5 to 2 ms.
-
Arrhythmia Analysis: Asystole, V Fib/V Tach, V Tach, VT>2 R on T, V Brady, Couplet, Bigeminy, Accelerated Ventricular arrhythmia, Multifocal PVCs, A Fib, Missing beat, Pause, Tachy, Brady, Trigeminy.
-
ST Segment Analysis:
-
Numeric Range: -9 to +9 mm (-0.9 to +0.9 mV).
-
Accuracy: +/- 0.2 mm or +/- 10%, whichever is greater, within the measurement range of -8 to 8 mm.
-
Numeric Resolution: 0.1 mm (0.01 mV).
-
ST Trends: Up to 168 h.
-
Impedance Respiration Range:
-
Adult/Pediatric: 4 to 120 resp/min.
-
Neonate: 4 to 180 resp/min.
-
Impedance Respiration Accuracy: +/- 5% or +/- 5 resp/min, whichever is greater.
-
Gain Range: 0.1 to 5 cm / Ohm.
-
TruSignal SpO2:
-
Pulse Oximetry Measurement Range: 1 to 100%.
-
Pulse Rate Measurement Range: 30 to 250 bpm.
-
Saturation Measurement Accuracy:
-
Without Motion-Adult/Pediatric Finger Sensor: 70 to 100% +/- 2%.
-
Without Motion-Neonate: 70 to 100% +/- 3%.
-
With Motion-Adult/Pediatric/Neonate: 70 to 100% +/-3%.
-
Low Perfusion-Adult/Pediatric: 70 to 100% +/- 3% (1 ~ 69% unspecified).
-
Pulse Rate Without Motion: +/- 2 bpm (Adult/Pediatric/Neonatal).
-
Nellcor OxiMax SpO2:
-
Pulse Oximetry Measurement Range: 1 to 100%.
-
Pulse Rate Measurement Range: 20 to 250 bpm.
-
Saturation Measurement Accuracy:
-
Adult: 70 to 100% +/- 2%.
-
Neo: 70 to 100% +/- 3%.
-
Low Perfusion: 70 to 100% +/-2 % (0~69% unspecified).
-
Pulse Rate Without Motion: +/- 3 bpm.
-
Pulse Rate With Motion: +/- 5 bpm.
-
NIBP:
-
Measurement Technique: Oscillometric with step deflation.
-
Modes: Manual, Automatic, Stat.
-
NIBP Measurement Ranges:
-
Systolic: Adult/Pediatric: 30 to 290 mmHg. Neonate: 30 to 140 mmHg.
-
MAP: Adult/Pediatric: 20 to 260 mmHg. Neonate: 20 to 125 mmHg.
-
Diastolic: Adult/Pediatric: 10 to 220 mmHg. Neonate: 10 to 110 mmHg.
-
Default Initial Inflation Pressure: Adult/Pediatric: 135 +/- 15 mmHg. Neonate: 10 to 110 mmHg.
-
Maximum Determination Rate: Adult/Pediatric: 2 min. Neonate: 85 s.
-
Overall Pressure Monitor: Adult/Pediatric: 294 +/- 6 to 330 mmHg. Neonate: 147 +/- 3 to 165 mmHg.
-
Invasive Blood Pressure:
-
Measurement Range: -40 to 320 mmHg (-5.3 to 42.7 kPa).
-
Measurement Accuracy: +/- 5% or +/- 2 mmHg, whichever is greater.
-
Frequency Response: 4 to 22 Hz.
-
Transducer Sensitivity: 5 ╬╝V/V/mmHg.
-
Temperature:
-
Numerical Display: T1, T2, T2-T1.
-
Measurement Range: 10 to 45 degrees C (50 to 113 degrees F).
-
Measurement Accuracy: +/- 0.1 degrees C without probe.
-
Display Resolution: 0.1 degrees C.
-
Probe: YSI probes recommended by GE Healthcare.
-
Networking:
-
Compatibility: CARESCAPE Network.
-
Wi-Fi Connectivity: IEEE 802.11a/b/g/n.
-
Wi-Fi Security: WPA-Personal; WPA2-Personal; WPA-Enterprise; WPA2-Enterprise.
-
Connectivity to EMR: HL7 outbound protocol direct through the CARESCAPE Gateway.
-
Power:
-
AC Input: 100 to 240 V +/- 10%, 50/60 Hz, 150 VA.
-
Protection: Class 1.
-
Battery: Exchangeable lithium-ion, 1 pcs max.
-
Charging time: Less than 4 hours to 90% capacity.
-
Run Time: Greater than 3 hours.
-
Dimensions: 270 mm x 290 mm x 150 mm (10.6" x 11.4" x 5.9").
-
Weight: 3.9 kg (8.6 lb) with battery.
-
Ingress Protection: IPX1.
Configurations
GE B105 Patient Monitor Configurations
-
#2099000-001-565659: GE Ohmeda SpO2.
-
#2099000-001-565662: Nellcor SpO2.
-
#2099000-001-565663: Masimo SpO2.
-
#2099000-001-565665: GE Ohmeda SpO2, IBP (Invasive Accessories Not Included).
-
#2099000-001-565660: Nellcor SpO2, IBP (Invasive Accessories Not Included).
-
#2099000-001-565661: Masimo SpO2, IBP (Invasive Accessories Not Included).
Literature
GE B105 Patient Monitor Literature
GE B105 Patient Monitor Brochure
Customer Reviews 0 out of 5 - (Total 0 ratings)